Test 3. Polarization Dependence - Agilent Technologies Agilent 86120B User Manual

Multi-wavelength meter
Table of Contents

Advertisement

Description
• 1310 nm and 1550 nm DFB lasers
• Optical attenuator
• Agilent 11896A Polarization Controller
C A U T I O N
Procedure
1 Turn on the lasers and allow them to warm up.
2 Set the polarization controller to a scan rate of 5.
3 On the Agilent 86120B, press the Preset key.
4 Connect the laser's optical output to the polarization controller's optical input.
5 Connect the polarization controller's optical output to the Agilent 86120B
6 Set the polarization controller to autoscan.
7 On the Agilent 86120B, press Peak WL, Appl's, and then DRIFT. Press MAX-MIN so
8 Wait five minutes, read the peak amplitude drift from the Agilent 86120B, and

Test 3. Polarization Dependence

Polarization Dependence is verified using the following devices:
Do not exceed +18 dBm source power. The Agilent 86120B's input circuitry can
be damaged when total input power exceeds 18 dBm.
Perform the following procedure first using the 1310 nm laser and then repeat
the steps using the 1550 nm laser.
being tested.
that both MAX and MIN in the softkey label are highlighted. The display shows
the total drift since the drift measurement was started.
compare with the specification listed in
Regulatory
Information".
Test 3. Polarization Dependence
Chapter 7, "Specifications and
Performance Tests
6-5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents