• NORMAL update mode unless noted.
• Configured to measure narrowband devices. Specifications do not apply when
Wavelength
Range
Absolute accuracy (lines separated by
Differential accuracy (characteristic)
Minimum resolvable separation
Display resolution
a. Signals of equal amplitude. For lines that are separated by less than 30 GHz, wavelength accuracy is reduced.
Specifications and Regulatory Information

Specifications

Specifications
Each laser line is assumed to have a linewidth (including modulation side-
bands) of less than 10 GHz.
All specifications apply when the instrument is in the following modes:
the instrument is configured to measure broadband devices.
ing broadband devices and chirped lasers" on page
30 GHz)
a
(characteristic)
7-6
Refer to "Measurement rate" on page
2-10.
700-1650 nm (182-428 THz)
±
±
3 ppm (
0.005 nm at 1550 nm,
±
2 ppm
20 GHz (0.16 nm at 1550 nm, 0.11 nm at 1300 nm)
0.001 nm, normal update mode
0.01 nm, fast update mode
2-14.
Refer to "Measur-
±
0.004 nm at 1310 nm)