Test 2. Sensitivity - Agilent Technologies Agilent 86120B User Manual

Multi-wavelength meter
Table of Contents

Advertisement

Description
• Optical power meter
• Optical attenuator
• 1310 nm and 1550 nm lasers (>0 dBm output power)
C A U T I O N
Procedure
1 Connect the laser's output to the optical attenuator's input.
2 Connect the optical attenuator's output to the optical power meter.
3 Adjust the attenuator for a reading of 0 dBm on the power meter. Record the
4 Adjust the attenuator for a reading of –35 dBm on the power meter. Record the
5 Disconnect the fiber-optic cable at the power meter's input and connect the
6 Reset the optical attenuator to the setting recorded in
7 Read the power and wavelength measured on the Agilent 86120B, and
8 Reset the optical attenuator to the setting recorded in
9 Read the power and wavelength measured on the Agilent 86120B, and
Performance Tests

Test 2. Sensitivity

Test 2. Sensitivity
Sensitivity is verified using the following devices:
Do not exceed +18 dBm source power. The Agilent 86120B's input circuitry can
be damaged when total input power exceeds 18 dBm.
Perform the following procedure first using the 1310 nm laser and then repeat
the steps using the 1550 nm laser.
attenuator's setting.
Attenuation at 0 dBm: ____________
attenuator's setting.
Attenuation at –35 dBm: ____________
cable to the Agilent 86120B being tested.
compared them to the specifications listed in
Regulatory
Information".
compared them to the specifications listed in
Regulatory
Information".
6-4
Step
3.
Chapter 7, "Specifications and
Step
4.
Chapter 7, "Specifications and