Function description
4.2
Gas Supplies
The current measured values of the O
adjusted. The inlet pressure for O
If the pressure falls below the limit value, a visible and audible alarm is generated. The operator is
informed about the cause of the alarm by a plain text message on the display. If the adjusted O
concentration is 21% or 100%, only an advisory message (no alarm) is displayed if the non-added gas
type fails.
If the pressure exceeds the limit value, the safety venting mechanism is activated and the continuous flow
is switched off. The Babylog will continue to operate only when the pressure has decreased to a
permissible level.
If the AIR supply fails, the Babylog control switches to O
to AIR supply.
4.3
Airway Pressure Monitoring
Depending on the selected mode and parameter settings, a certain airway pressure time profile is set. If
the actual profile deviates from the set profile, the Babylog generates an alarm.
4.4
Disconnect Monitoring
Babylog recognizes whether or not the hose system has been connected correctly. If the hose system has
been connected incorrectly, the continuous flow is (partly) directed to atmosphere. No pressure builds up
in the hose system. Babylog checks during each breathing cycle whether a sufficiently high pressure is
available during inspiration. Since inspiratory breaths occur in the mandatory ventilation modes
IPPV/IMV, SIPPV and SIMV, this monitoring does not function in the CPAP mode.
Paw
Pinsp
Tdis
Pdis
PEEP
Tdis > 25 ms !
Pdis = Pinsp - PEEP
Fig. 4: Disconnect Monitoring
20
Dräger Medical AG & Co. KGaA
and AIR gas supplies are considered when the valve bank is
2
and AIR is 1.7 bar.
2
+ PEEP
4
6173.3
supply. If the O
supply fails, Babylog switches
2
2
time
Zeit
Babylog 8000
2