Dräger Babylog 8000 Technical & Service Manual page 68

Baby intensive care ventilator
Hide thumbs Also See for Babylog 8000:
Table of Contents

Advertisement

Babylog 8000
• Remove the battery 5 and install a new one.
5
Fig. 8: Replacing the Battery
• Re-assemble the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC. The battery for power failure
alarm is recharged automatically during operation.
After replacing the battery, allow the device to operate for at least half an hour in order to sufficiently
charge the battery.
No acoustic alarm will sound in the event of power failure if
the battery is not sufficiently charged. Risk of severe injury or
death. Allow the device to operate for at least half an before
clinical application.
66
Dräger Medical AG & Co. KGaA
6173.3
Replacing Non-Repairable Items