Dräger Babylog 8000 Technical & Service Manual page 76

Baby intensive care ventilator
Hide thumbs Also See for Babylog 8000:
Table of Contents

Advertisement

Babylog 8000
• Set the pressure reducer to 1.7
• Set the insp. flow adjuster to 30 L/min. The manometer should indicate at least 1.3 bar.
• Lock both adjusting rings by pressing them in direction of the arrows C.
• Make sure that the hose system is free of leaks.
Fig. 21: Locking the Pressure Reducer
• Re-assemble the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC and it return it to service ready-
for-use.
74
Dräger Medical AG & Co. KGaA
-0.1
bar.
C
C
6173.3
Replacing Non-Repairable Items

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Babylog 8000scBabylog 8000 plus

Table of Contents