12
Testing the Pressure Regulator
• Connect the inspiratory socket with the expiratory socket.
• Adjust the device as shown in the Figure below
Flow sensor cable
Fig. 7: Pressure Regulator
• Connect the device to the O
The following message is displayed on the device: "no medical air".
• Connect the device to the medical AIR supply only.
The following message is displayed on the device: "no oxygen".
6173.3 Babylog 8000/8000 plus/8000 SC
CPAP
30
15
supply only.
2
12/00
General
Page 13