Dräger Babylog 8000 Technical & Service Manual page 147

Baby intensive care ventilator
Hide thumbs Also See for Babylog 8000:
Table of Contents

Advertisement

14
Testing the Mixer
• Carry out an O
2
• Test set-up (see Figure below):
Flow sensor cable
Fig. 9: Mixer
• Select the following inspiratory flows from the "list" menu: 1.2 / 1.3 / 1.4 / 1.5 / 1.6
/ 1.7 / 1.8 / 1.9 / 2.0 / 4 / 5 / 6 / 8 / 9 / 9.9 / 11 / 12 / 13 / 14 / 15 / 30
Allow approx. 1 min between setting and measurement of flow values (< 2 L/min). If
no values is set, the alarm "FiO
15
Testing the Patient System Heating
• Put your hand against the expiratory valve block.
The temperature should exceed the room temperature.
6173.3 Babylog 8000/8000 plus/8000 SC
calibration.
60
var
deviation" is allowed to occur.
2
12/00
CPAP
5
30
General
Page 15