Dräger Babylog 8000 Technical & Service Manual page 95

Baby intensive care ventilator
Hide thumbs Also See for Babylog 8000:
Table of Contents

Advertisement

Error List
Table 1: Device error messages
Display error
Monitoring PCB
number
error number
505
506 to 513
800
801
802
803
804
805
806
807
809
Dräger Medical AG & Co. KGaA
Meaning
CPU PCB software error
CPU PCB software error.
FiO
too high by more than 10
2
vol.%.
FiO
too low by more than 10
2
vol.%.
FiO
measurement deviation
2
between master (CPU PCB and
I/O PCB) and Monitoring PCB is
too high.
Measured Pinsp to potentiometer
setting Pinsp too high by more
than 25 mbar (cmH
O).
2
Pinsp greater than 110 mbar
(cmH2O).
Pinsp less than -5 mbar (cmH2O)
Flow measurement on the Flow
PCB is faulty
Potentiometer evaluation is faulty:
Master (CPU PCB and
Frontcontroller PCB) and
Monitoring PCB read different
values.
Battery or battery charge faulty
(battery voltage < 7 V)
Babylog 8000