Replacing The Lip Seals Every 2 Years - Dräger Babylog 8000 Technical & Service Manual

Baby intensive care ventilator
Hide thumbs Also See for Babylog 8000:
Table of Contents

Advertisement

Babylog 8000
4

Replacing the Lip Seals Every 2 Years

• Lift the toggle 1 and take out the expiratory valve.
D
1
Fig. 4: Removing the Expiratory Valve
• Use, for example, a spherical-head 2.5 mm screwdriver to remove the old lip seals 2. Insert the new
lip seals and make sure they fit properly.
D
2
2
Fig. 5: Location of the Lip Seals
• Re-mount the expiratory valve. To do so, lift up the toggle and place the expiratory valve to the front
using the guide rods.
• Lock the expiratory valve with the toggle 1.
64