Recommended separation distances
Recommended separation distances between portable and mobile RF telecommunication
max.
3 V/m
P
(W)
distance*
EIRP
(m)
0.001
0.06
0.003
0.10
0.010
0.18
0.030
0.32
0.100
0.58
0.200
0.82
0.250
0.91
1.000
1.83
2.000
2.60
3.000
3.16
Information re separation distances (IEC 60601-1-2: 2001, tables 205 and 206)
* 3 V/m distance to transmitters with frequencies from 150 kHz to 2.5 GHz, otherwise 1 V/m distance.
devices and Babylog 8000 plus
1 V/m
distance*
(m)
0.17
0.30
0.55
0.95
1.73
2.46
2.75
5.48
e.g. GSM 1800- / GSM 1900- / UMTS- mobiles,
7.78
9.49
Page 4 of 4
Note
e.g. WLAN 5250 / 5775 (Europe)
e.g. WLAN 2440 (Europe), Bluetooth
e.g. WLAN 5250 (not in Europe)
e.g. DECT devices
WLAN 5600 (not in Europe)
e.g. GSM 900 mobiles