Replacing Non-Repairable Items
• Remove the fastening screws 9 from the pressure reducers.
• Pull out the pressure reducers.
9
9 9
Fig. 16: Supply Block
• Replace the pressure reducers 10 for AIR and O
12 (part number M23154).
10
11
Fig. 17: Pressure Reducers
• Mount the pressure reducers onto the supply block.
• Mount the supply block.
• Re-insert the gasket and the filters (wire fabric side shows towards inside).
• Mount the gas supply connection plate.
Dräger Medical AG & Co. KGaA
9
6173.3
including the O-rings 11 (part number D19080) and
2
Babylog 8000
12
71