Dräger Babylog 8000 Technical & Service Manual page 149

Baby intensive care ventilator
Hide thumbs Also See for Babylog 8000:
Table of Contents

Advertisement

19
Tests According to VDE 0751 Part 1
(Follow the national and international laws, standards and regulations
applicable in your country)
Power cable
Mains fuse-unit
(the specifications on the fuse-unit must match those on the rating plate)
Protective earth conductor test < 0.2 W
Equivalent device leakage current < 1.0 mA
Patient leakage current < 4.0 mA
20
Place fully functional unit at customer's disposal.
21
Tested
Date: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
6173.3 Babylog 8000/8000 plus/8000 SC
Name: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
12/00
General
Page 17