Set up the OXIMAX N-65
Connect a Nellcor O
SpO
sensor port. You may use an extension cable to
2
provide more distance between the monitor and the
sensor. Use only the DEC-4 extension cable available
from Nellcor.
Inaccurate Monitor Measurements
Physiological conditions, medical procedures, or external
agents that may interfere with the monitor's ability to
detect and display accurate measurements include:
•
incorrect application of the O
•
placement of the O
with a blood pressure cuff, arterial catheter, or
intravascular line
•
excessive patient activity
•
intravascular dyes, such as indocyanine green or
methylene blue
•
externally applied coloring agents, such as nail
polish, dye, or pigmented cream
28
M
SpO
sensor to the monitor
XI
AX
2
1
SpO
O
M
Sensor Port
—
XI
AX
2
M
XI
M
sensor on an extremity
XI
AX
O
Max N-65 Operator's Manual