Set Alarm Silence Duration

Discussion

When the Alarm Silence Duration display shows, you
can adjust the alarm silence duration of the high,
medium, and low-priority alarms.
A low battery low-priority alarm cannot be silenced.

Procedure

To set the alarm silence duration:
1. From the main monitor screen, press the Alarm
Silence button until SEC or OFF appears on the
pulse rate area of the display. Then, release the
button.
O
Max N-65 Operator's Manual
XI
OXIMAX N-65 Operation
Note: When the monitor times-out in 30
seconds, the parameter is set and the monitor
returns to the normal mode.
WARNING: Do not silence the audible
alarm function, or decrease the audible
alarm volume, if patient safety could be
compromised.