Nellcor OxiMax N-65 Operator's Manual page 127

Hendheld pulse oximeter
Hide thumbs Also See for OxiMax N-65:
Table of Contents

Advertisement

Audible
Indicator
Alarm Volume
Setting
Beep Volume
setting
O
Max N-65 Operator's Manual
XI
Audible Indicators
Parameter
Volume level
Adjustable, 40 to 52
dB(A), at one meter
Pitch (±30 Hz)
752 Hz
On pulse width
500 msec
(±20 msec)
Off Interval (±20 msec)
10 msec
Number of pulses in
1
burst
Repetition Pause
N/A
(±2 sec.)
Priority
1
Volume level
Adjustable, 42 to 52
dB(A), at one meter
Pitch (±30 Hz)
1500 Hz
On pulse width
500 msec
(±20 msec)
Off Interval (±20 msec)
10 msec
Number of pulses in
1
burst
Repetition Pause
N/A
(±2 sec.)
Priority
2
Specifications
Value
121

Hide quick links: