Nellcor OxiMax N-65 Operator's Manual page 100

Hendheld pulse oximeter
Hide thumbs Also See for OxiMax N-65:
Table of Contents

Advertisement

Troubleshooting
See Table 5, "Error Codes", on page 94 for a list of error
codes and corrective actions.
When an error occurs, the unit:
stops monitoring
removes all information from the display and
displays the message "Er XXX"
sounds a low priority alarm
When an error occurs, you must:
Turn the monitor off.
Wait 10 seconds and turn the monitor on.
If the error appears, follow the action(s) listed in
Table 5.
Error
Code
10
11
17
19
273
274
275
276
94
Table 5: Error Codes
Action
Check/replace sensor/extension cable.
1 —
Replace batteries.
2 —
Notify service personnel.
Check/replace sensor/extension cable.
Check/replace sensor/extension cable.
1 —
Restart the monitor.
2 — Set the time and date.
3 —
Notify service personnel.
Return the monitor for reprogramming.
Check/replace sensor/extension cable.
Replace with O
M
sensor.
XI
AX
O
Max N-65 Operator's Manual