Table 12: Electromagnetic Immunity
IEC 60601
Immunity Test
Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Recommended Separation
Distance
3.5
------ -
d
=
P
V 1
3.5
------ -
d
=
P
E 1
80 MHz to
800 MHz
O
Max N-65 Operator's Manual
XI
Electromagnetic
Compliance
Environment -
Level
Guidance
3 Vrms
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the O
3 Vrms
65, including the cables,
than the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
a
electromagnetic site survey
less than the compliance level in each
b
frequency range.
Specifications
M
N-
XI
AX
, should be
135