Oximax N-65 Operation; Introduction - Nellcor OxiMax N-65 Operator's Manual

Hendheld pulse oximeter
Hide thumbs Also See for OxiMax N-65:
Table of Contents

Advertisement

O
M
N - 6 5 O p e r a t i o n
X I
A X

Introduction

The parameters of the O
according to Table 2. The parameters may be set on an
individual basis by the clinician.
Note: This section contains detailed information on
setting parameters. For quick reference information, see
Set Menu Parameters on page 108 and Set Time/Date
Parameters on page 110.
Parameters remain in effect until the monitor is turned
off.
Table 2: Parameter Ranges and Defaults
Parameter
SpO
%
Alarm Limit
SpO
%
Alarm Limit
Pulse Rate Upper
Alarm Limit
Pulse Rate Lower
Alarm Limit
O
Max N-65 Operator's Manual
XI
M
N-65 are set at the factory
XI
AX
Ranges/
Selections
Lower Alarm
Upper
2
Limit plus 1 to
100%
20% to Upper
Lower
2
Alarm Limit
minus 1
Lower Alarm
Limit plus 1 to
250 bpm
30 bpm to Upper
Alarm Limit
minus 1
Factory
Defaults
100%
85%
170 bpm
40 bpm
35

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents