After Taking Measurements; Sensor Disconnected; Sensor Off; Automatic Shutdown - Nellcor N-395 Operator's Manual

Table of Contents

Advertisement

After Taking Measurements
If a pulse was previously acquired and then lost, the N-395
enters Pulse Search, and the Pulse Search indicator lights. The
last detected readings are displayed while the monitor searches
for a valid pulse. When the monitor considers the pulse "lost,"
it displays zeroes and a high priority alarm sounds.
When a valid pulse is detected, the N-395 exits the Pulse Search
Mode and displays the current readings. The Pulse Search
indicator goes out.
Sensor Disconnected
If the sensor cable becomes disconnected from the sensor or the
monitor during monitoring, a low priority alarm sounds, values
for SpOz and pulse rate are replaced with dashes, and SENSOR
DISCONNECTED is displayed on the screen.
Sensor Off
If the sensor becomes disconnected froin the patient during
monitoring, a low priority alarm sounds, values for SpOz and
pulse rate are replaced with dashes, and SENSOR OFF is
displayed on the screen.
Automatic Shutdown
When all of the following conditions are present for 15 minutes,
the N-395 will automatically shut down:
Running on battery power
No buttons have been pressed
No pulse has been detected (for example, when no patient
is connected to the sensor or the sensor is disconnected)
No alarms are present (other than low battery or a non-
correctable error)