6. Press the Up Arrow or Down Arrow buttons to
increase or decrease the lower limit.
7. Press the Menu button to set the limit value. The
BPM upper alarm limit displays.
The pulse rate upper alarm limit range is 31 to 250.
The lower value of the pulse rate upper alarm limit is
limited to one number above the pulse rate low alarm
limit. The pulse rate upper alarm limit cannot be set
equal to or lower than the pulse rate low alarm limit.
O
Max N-65 Operator's Manual
XI
OXIMAX N-65 Operation
59