Fluke 917X Series Technical Manual page 98

Hart metrology wells
Table of Contents

Advertisement

917X Series Metrology Wells
Non-SCPI Command by Function or Group
lf [n]
Read or set the serial interface linefeed enable, where "n" is a value 1 or 0. [0] = LF
OFF, [1] = LF ON. The default setting is Off. (Off and on may be used in place of 0
and 1 respectively). If "n" is left blank, the command will be treated as a query. This
query responds with the command string and a Colon followed by the LF setting.
Refer to SYST:COMM:SER:LIN
Read Example: lf
lf: OFF
Set Example: lf on
s [n]
Read or Set the temperature control set-point in °C or °F (based on current system
units). Where "n" is a real value with acceptance limits based on the model. If "n" is
left blank, the command will be treated as a query. This query responds with the com-
mand string "set:" followed by the temperature setting and associated units. Refer to
SOUR[1]:SPO command
Read Example: s
set: 100.00 C
Set Example: s 250
sa [n]
Read or Set the serial interface auto printing interval. Where "n" is an integer value
from 0-60. If "n" is 0, the auto print will be disabled. Values range from 1 to 60 and
are in seconds. If "n" is left blank, the command will be treated as a query. This query
responds with the command string "sa" and a Colon followed by the interval setting.
Refer to SYST:COMM:SER:TIME command
Read Example: sa
sa: 5
Set Example: s 10
t
Read the control temperature in °C or °F (based on current system units). This com-
mand is query only and responds with the command string and a Colon followed by
the temperature and associated units. Refer to SOUR[1]:SENS:DAT command
Read Example: t
t: 99.988 C
90