SOUR2:LCON:PBAN[?] [n]
NOTE: This command is unconditionally protected and requires a password to
set it.
Read or set the gradient temperature control loop proportional band, °C
Read Example: SOUR2:LCON:PBAN?
Response: 5.0
Set Example: SOUR2:LCON:PBAN 10
The gradient proportional band is the gain in °C that the unit's PID controller uses for
gradient control.
SOUR2:SENS:CAL:PAR[n][?] [n]
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set a gradient temperature calibration parameter (:PAR1, :PAR2, :PAR3)
Read Example: SOUR2:SENS:CAL:PAR1?
Response: 0.0
Set Example: SOUR2:SENS:CAL:PAR1 5.0
Parameter 1 is the offset in °C for the axial gradient at the 1st gradient calibration
point (GRAD 1).
Parameter 2 is the offset in °C for the axial gradient at the 2nd gradient calibration
point (GRAD 2)
Parameter 3 is the offset in °C for the axial gradient at the 3rd gradient calibration
point (GRAD 3). (Model Dependent)
SYST:BEEP:IMM
Beep the system beeper
Example: SYST:BEEP:IMM
The system beeper should make an audible sound in response to this command.
SYST:COMM:SER:BAUD[?] [<baud>]
Read or set serial interface baud rate
Read Example: SYST:COMM:SER:BAUD?
Response: 2400
Set Example: SYST:COMM:SER:BAUD 9600
Digital communication interface
Serial Commands - Alphabetic Listing
85