Fluke 917X Series Technical Manual page 84

Hart metrology wells
Table of Contents

Advertisement

917X Series Metrology Wells
Serial Commands - Alphabetic Listing
This command returns the temperature when the switch changed state. The units of the
temperature returned is determined by the units on the display.
OUTP:SLAV:RAT[?] [n]
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set the slave heater ratio
Read Example: OUTP:SLAVE:RAT?
Response: 1.0
Set Example: OUTP:SLAVE:RAT 2.0
The heater ratio is the ratio in percentage of full power between the main heaters and
the bottom heaters in the hot units. For more information regarding this parameter,
refer to the Controller Operation section.
OUTP[:STAT][?] [0|1]
NOTE: Depending on group password setting, this command is conditionally
protected.
Read or set the heat output enable, off (0) or on (1)
Read Example: OUTP:STAT?
Response: 0
Set Example: OUTP:STAT 1
This command is password protected via group 1(Set Point protection).
This command reads the active heating or cooling output status. A "0" is returned if
the output status is off, and a "1" is returned if the output status is on.
OUTP[1]:DAT?
Read the main heat output percent
Example: OUTP[1]:DAT?
Response: 18.0
This command returns the current main zone heater duty cycle.
OUTP2:DAT?
Read the gradient heat output percent
Example: OUTP2:DAT?
76