Serial Commands - Alphabetic Listing
A value of 0 is returned if the cutout set point has not been reached. Otherwise a value
of 1 is returned and the cutout set point has been reached.
SOUR[1]:RATE?
Read the control temperature rate of change, °C or °F per minute
Example: SOUR:RATE?
Response: 0.531
The response to this command starts out high initially and decreases as the set point is
reached.
SOUR:SENS:CAL:PAR[n][?] [n]
NOTE: This command is unconditionally protected and requires a password to
set it.
Read or set a control temperature calibration parameter (:PAR1, :PAR2, :PAR3)
Read Example: SOUR:SENS:CAL:PAR1?
Response: 0.0
Set Example: SOUR:SENS:CAL:PAR2 0.02
This command reads or sets the calibration parameters for main control. See the "Cali-
bration of your Metrology Well" section for more detail.
SOUR[1]:SENS:DAT? [TEMP]
Read the control temperature, °C or °F
Example: SOUR:SENS:DAT?
Response: 30.285°C (current control temp)
The current control temperature is returned if the above example is used or if TEMP is
appended to the end of the example.
SOUR[1]:SENS:DAT? [RES]
Read the control sensor resistance
Example: SOUR:SENS:DAT? RES
Response: 111.28
When RES is appended to the end of the example above, the internal sensor resistance
is returned.