Fluke 917X Series Technical Manual page 92

Hart metrology wells
Table of Contents

Advertisement

917X Series Metrology Wells
Serial Commands - Alphabetic Listing
Example: SOUR:STAB:DAT?
Response: 0.306
The controller stability is returned.
SOUR[1]:STAB:LIM[?] [n]
NOTE: Depending on group password setting, this command is conditionally
protected.
Read or set the control temperature stability limit, °C or °F
Read Example: SOUR:STAB:LIM?
Response: 0.005
Set Example: SOUR:STAB:LIM .005
This command is password protected via group 5(System protection).
SOUR[1]:STAB:TEST?
Read the temperature stability status
Example: SOUR:STAB:TEST?
Response: 0
A value of 0 is returned if the controller is not stable at the current set point.
Otherwise a value of 1 is returned if the controller is stable at the current set point.
SOUR2:LCON:INT[?] [n]
NOTE: This command is unconditionally protected and requires a password to
set it.
Read or set the gradient temperature control loop integral time, seconds
Read Example: SOUR2:LCON:INT?
Response: 40.0
Set Example: SOUR2:LCON:INT 20
The gradient integral is the integration time in seconds that the unit's PID controller
uses for gradient control.