Digital communication interface
Serial Commands - Alphabetic Listing
Parameters for two probes can be stored in the instrument. This command selects the
probe number.
CALC:CONV:SNUM[?] [n]
Read or set reference probe serial number.
Read Example: CALC:CONV:SNUM?
Response: 1234
Set Example: CALC:CONV:SNUM 1560D
This command allows the user to enter a reference probe serial number.
CALC:CONV:TEST[?] [n]
Test the reference probe resistance to temperature calculation
Read Example: CALC:CONV:TEST? 100.000
Response: 0.0100
This is a read only command and tests the external probe resistance.
DISP:LANG[?] [n]
Read or set the display language.
Read Example: DISP:LANG?
Response: 0
Set Example: DISP:LANG 1
This command is used to read or set the Display Language. A value of 0, 1, or 2 sets
the Display Language to English, French, or Chinese respectively.
INP:SWIT:CLOS?
Read the switch input state, open (0) or closed (1)
Example: INP:SWIT:CLOS?
Response: 0
This command returns a 0 if the switch is open and a 1 if the switch is closed.
INP:SWIT:TEMP?
Read the temperature when the switch changed state.
Example: INP:SWIT:TEMP?
Response: 105.0