Controller operation
Main Menu
TEMP 1
The TEMP 1 parameter is the offset in °C for the display accuracy at the 1st calibra-
tion point.
TEMP 2
The TEMP 2 parameter is the offset in °C for the display accuracy at the 2nd calibra-
tion point.
TEMP 3
The TEMP 3 parameter is the offset in °C for the display accuracy at the 3rd calibra-
tion point.
GRAD 1
The GRAD 1 parameter is the offset in °C for the axial gradient at the 1st gradient
calibration point.
GRAD 2
The GRAD 2 parameter is the offset in °C for the axial gradient at the 2nd gradient
calibration point.
GRAD 3 (Model Dependent)
The GRAD 3 parameter is the offset in °C for the axial gradient at the 3rd gradient
calibration point.
DAY
The DAY parameter is the day on which the instrument was calibrated.
MONTH
The MONTH parameter is the month in which the instrument was calibrated.
YEAR
The YEAR parameter is the year in which the instrument was calibrated.