917X Series Metrology Wells
Serial Commands - Alphabetic Listing
In the above, example the sequence program set-point 1 is returned. If PROG:NUMB
PPO? is entered, the return value is the current set-point number of the sequence
program. PSO refers to the amount of time the program soaks at the given set point
when stable. PCYC refers to the number of cycles the program runs. PDIR refers to
the direction the program takes. A PDIR response of 0 refers to a forward path only.
A response of 1 refers to a forward path and a return path. The following suffixes are
read only PPO, PSO, PCYC, PDIR. The other suffixes are read and write.
PROG:STAT[?] [RUN|STOP|CONT]
NOTE: Depending on group password setting, this command is conditionally
protected.
Read or set the selected program execution state
Read Example: PROG:STAT?
Response: 0
Set Example: PROG:STAT STOP
This command is password protected via group 4 (program protection).
If a Sequence program is not running then a value of 0 is returned otherwise a 1 is
returned.
READ?, MEAS? or FETC?
Read the reference thermometer input temperature, C or F.
Example: READ?
Response: 264.262
If the external reference probe is enabled, the reference temperature is returned other-
wise 0.0 is returned.
SENS:CAL:PAR[n][?] [n]
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set a reference input calibration parameter (:PAR1, :PAR2, :PAR3)
Read Example: SENS:CAL:PAR1?
Response: 0.2
Set Example: SENS:CAL:PAR2 0.092
Reference thermometer input commands to verify or set TEMP 1 (PAR1), TEMP 2
(PAR2), or TEMP 3 (PAR3) calibration parameters.