<month> indicates a one or two digit number is required.
<day> indicates a one or two digit number is required.
<hour> indicates a one or two digit number is required.
<minute> indicates a one or two digit number is required.
<second> indicates a one or two digit number is required.
<baud> indicates a valid baud number is required.
*CLS
Clear the status registers
Example: *CLS
This command has no response.
Clears all status registers(events, operations etc).
*IDN?
Read the product information (Manufacturer, Model Number, Serial Number, and
Firmware Version)
Example: *IDN?
Response: HART,9170,0,1.00
*OPT?
Read the product configuration, reference hardware enabled (1) or not (0) (see
SYST:CONF:REF)
Example: *OPT?
Response: 1
This command is a read only command and returns the state of the reference function-
ality (0, 1).
CAL:DAT:CAL[?] [<year>,<month>,<day>]
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set the calibration date
Read Example: CAL:DAT:CAL?
Response: 2005,1,1
Set Example: CAL:DAT:CAL 2006,12, 30
This command reads or sets the calibration date for the unit.
Digital communication interface
Serial Commands - Alphabetic Listing
73