SENS:DAT?
Read the reference input resistance
Example: SENS:DAT?
Response: 199.9366
This command returns the resistance in ohms of the reference probe.
SENS:STAT?
Read or set the reference probe enable state off (0) or on (1).
Read Example: SENS:STAT?
Response: 1
Set Example: SENS:STAT 1
The Reference Thermometer function of the instrument can be enabled or disabled
through this command.
SOUR[1]:LCON:DER[?] [n]
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set the main control loop derivative time, seconds
Read Example: SOUR1:LCON:DER?
Response: 1.5
Set Example: SOUR1:LCON:DER 5
The main zone derivative is the derivative time in seconds that the unit's PID control-
ler used for main zone control.
SOUR[1]:LCON:INT[?] [n]
NOTE: This command is unconditionally protected, which requires a password
to set it.
Read or set the main control loop integral time, seconds
Read Example: SOUR1:LCON:INT?
Response: 20.0
Set Example: SOUR1:LCON:INT 10
The main zone integral is the integration time in seconds that the unit's PID controller
uses for main zone control.