Fluke 917X Series Technical Manual page 82

Hart metrology wells
Table of Contents

Advertisement

917X Series Metrology Wells
Serial Commands - Alphabetic Listing
CALC:CONV:NAME[?] [CVD|ITS]
NOTE: Depending on group password setting, this command is conditionally
protected.
Read or set the reference probe type
Read Example: CALC:CONV:NAME?
Response: CVD
Set Example: CALC:CONV:NAME I90
This command is password protected via group 3(Reference Protection).
CALC:CONV:PAR:CAT?
Read the list of reference probe characterization parameter names
Example: CALC:CONV:PAR:CAT?
Response: "RTPW","A","B","C","A4","B4"
This command is a read only command, which returns the available parameters for the
current probe type.
CALC:CONV:PAR:VAL[?] par[,<n>]
NOTE: Depending on group password setting, this command is conditionally
protected.
Read or set a reference probe characterization parameter (A, A4, AL (Alpha), B, B4,
BE (Beta), C, DE (Delta), RTPW)
Read Example: CALC:CONV:PAR:VAL? RTPW
Response: 100.000
Set Example: CALC:CONV:PAR:VAL A, 0.00385
This command is password protected via group 3 (Reference Protection).
This command returns an ITS-90 reference probe characterization parameter as de-
sired by the user.
CALC:CONV:SEL[?] [n]
Read or set the reference probe characterization parameter set.
Read Example: CALC:CONV:SEL?
Response: 1
Set Example: CALC:CONV:SEL 1