Fluke 917X Series Technical Manual page 89

Hart metrology wells
Table of Contents

Advertisement

Digital communication interface
Serial Commands - Alphabetic Listing
Read Example: SOUR:PROT? MAX
Response: 200
Set Example: SOUR:PROT 200
Returns the maximum calibrated cutout temperature.
SOUR[1]:PROT:CLE
Reset the temperature cutout
Example: SOUR:PROT:CLE
This command has no response.
If the Metrology Well exceeds the temperature set in the soft cutout menu (or when
using the command SOUR:PROT) or if it exceeds the maximum operating tempera-
ture of the instrument, a cutout condition occurs. If this happens, the unit enters cutout
mode and will not actively heat or cool until the user issues this command to clear the
cutout.
SOUR[1]:PROT:HIGH:BEEP[?] [n]
Reads or sets the cutout alarm.
Read Example: SOUR:PROT:HIGH:BEEP?
Response: 0
Set Example: SOUR:PROT:HIGH:BEEP 1
This command enables or disables the Cutout Alarm. 0 disables the Alarm and 1 en-
ables the Alarm.
SOUR[1]:PROT:SOFT[?] [0|1]
Read or set the soft cut-out enable, off (0) or on (1)
Read Example: SOUR:PROT:SOFT?
Response: 1
Set Example: SOUR:PROT:SOFT 0
If this command is issued, as in the above set example, the hard cutout limit is used,
otherwise the soft cutout is used.
SOUR[1]:PROT:TRIP?
Read the temperature cutout tripped state
Example: SOUR:PROT:TRIP?
Response: 0
81