917X Series Metrology Wells
Serial Commands - Alphabetic Listing
SOUR[1]:LCON:PBAN[?] [n]
NOTE: This command is unconditionally protected and requires a password to
set it.
Read or set the main control loop proportional band, °C
Read Example: SOUR1:LCON:PBAN?
Response: 1.5
Set Example: SOUR1:LCON:PBAN 7
The main zone proportional band is the gain in °C that the unit's proportional-integral-
derivative (PID) controller uses for main zone control.
SOUR[1]:LIST:SPO[i][?] [n]
NOTE: Depending on group password setting, this command is conditionally
protected.
Read or set a main temperature preset set-point
Read Example: SOUR1:LIST:SPO6?
Response: 25.00
Set Example: SOUR1:LIST:SPO6 100.00
This command is password protected via group 1(Set Point protection).
This sets the preset set-points found in Temp Menu under Setup.
SOUR[1]:PROT[?] [n]
NOTE: Depending on group password setting, this command is conditionally
protected.
This command is password protected via group 2(Cutout Protection).
Read or set the temperature cutout set-point in C or F
Read Example: SOUR:PROT?
Response: 140
Set Example: SOUR:PROT 140.00
Returns the current value of the cutout set-point.
SOUR[1]:PROT? MAX
Read or set the temperature cutout limit
80