Fluke 917X Series Technical Manual page 26

Hart metrology wells
Table of Contents

Advertisement

917X Series Metrology Wells
Reference Probe (-R models only)
4. Select the calibration type for the probe (ITS-90, CVD, or Res) using the 4 or
5. Use the numeric keys to enter the serial number and calibration coefficients of
6. After the reference probe has been enabled, the focus of the temperature display
18
5 keys. Press "ENTER" to accept the calibration type.
the probe. Press "ENTER" after each parameter is entered to accept the new
parameter value. Calibration coefficients can be found on a report of calibration
that was shipped with your probe. If no coefficients can be found, contact
the manufacturer or vendor of the probe for assistance. If the probe is out of
calibration, Fluke Calibration offers calibration services. Contact an Authorized
Service Center for assistance on obtaining a calibration for the probe.
can be set. The focus is disabled if the reference probe is disabled. From the
reference menu, press the "Focus" soft key (Main Menu: REF MENU|FOCUS).
Use the 4 or 5 key to select reference focus. Press "ENTER" to accept the
selection. The largest temperature on the display will now be the reference
temperature.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

9170917191729173

Table of Contents