SOUR[1]:SLEW[?] [n]
NOTE: Depending on group password setting, this command is conditionally
protected.
Read or set the control set-point slew rate, °C/F per minute
Read Example: SOUR:SLEW?
Response: 500
Set Example: SOUR:SLEW 100
This command is password protected via group 1(Set Point protection).
This command sets the controller ramp rate (°C/F per min).
SOUR[1]:SPO[?] [n]
NOTE: Depending on group password setting, this command is conditionally
protected.
Read or set the control set-point, °C or °F
Read Example: SOUR:SPO?
Response: 50.000
Set Example: SOUR:SPO 100.00
This command is password protected via group 1(Set Point protection).
This command returns the value of the control set point based on the system tempera-
ture unit.
SOUR[1]:STAB:BEEP[?] [0|1]
NOTE: Depending on group password setting, this command is conditionally
protected.
Read or set the stability alert enable
Read Example: SOUR:STAB:BEEP?
Response: 1
Set Example: SOUR:STAB:BEEP 0
Enable or disable the audible stability alert.
SOUR[1]:STAB:DAT?
Read the control temperature stability, °C or °F
Digital communication interface
Serial Commands - Alphabetic Listing
83