Digital communication interface
Serial Commands - Alphabetic Listing
A response of 0 implies that a period will be used for a decimal. Otherwise, a comma
will be used as the decimal.
SYST:ERR?
Read the errors from the error queue
Example: SYST:ERR?
Response: command protected
This command response reports the errors in the error queue.
SYST:FAN? [n]
Read or set the system fan mode.
Example: SYST:FAN?
Response: 2
Set Example: SYST:FAN 1
The system fan has three operating modes: Auto (2), Medium (1), low (0).
SYST:PASS:CDIS
Disable access to password protected setting commands
Example: SYST:PASS:CDIS
This command has no response.
This command disables the system password protection.
SYST:PASS:CEN [n]
Enable access to password protected setting commands
Example: SYST:PASS:CEN 1234
This command has no response.
This command enables the system password. This password needs to be enabled in
order to use the group conditionally protected commands. When the power of the
instrument is cycled, system password protection is disabled.
SYST:PASS:CEN:STAT?
Read the access state of password protected setting commands
Example: SYST:PASS:CEN:STAT?
Response: 0
This command reports the current status of the system password.