3
Tuning
Generating a Detailed Tune Report
Generating a Detailed Tune Report
6475A and Ultivo
A detailed tune report can be generated after autotune or checktune has run. Only
the latest tune report will generate a detailed report; e.g. if the last tune run was a
checktune, then only a detailed checktune report will be generated.
1 In the Method Editor window, click the TQ tab.
2 Click Tune > Autotune in the left pane.
3 Click
This button locks control of the TQ instrument. When tune has control of the
LC/TQ instrument, a single sample run or a worklist cannot be started.
4 Click
run either a checktune or an autotune.
5 Click
instrument.
Agilent Triple Quadrupole LC/MS System User Guide
Request tune control.
Generate Detailed Tune Report. If this button is not available, then
Release tune control in the toolbar to release control of the TQ
50