HP 1660 Series Service Manual page 68

Logic analyzers
Hide thumbs Also See for 1660 Series:
Table of Contents

Advertisement

Connect the HP 1662A or HP 1663A Logic Analyzer to the Pulse Generator
Testing
Connect to
Combination
HP 8131A
Channel 1 Output
1
Pod 1, channel 3
Pod 2, channel 3
Pod 3, channel 3
Pod 4, channel 3
Activate the data channels that are connected according to one of the previous
3 3
tables.
a a Press the Format key.
b b Select the field showing the channel assignments for one of the pods being tested.
Press the Clear entry key. Using the arrow keys, move the selector to the data
channels to be tested, then press the Select key. An asterisk means that a channel is
turned on. When all the correct channels of the pod are turned on, press the Done
key. Follow this step for the remaining pods.
To test the multiple-clock, multiple-edge, state acquisition (logic analyzer)
Connect to
HP 8131A
Channel 1 Output
Pod 1, channel 11
Pod 2, channel 11
Pod 3, channel 11
Pod 4, channel 11
Connect to
Connect to
HP 8131A
HP 8131A
Channel 2 Output
Channel 2 Output
J-clock
K-clock
L-clock
M-clock
3–39