HP 1660 Series Service Manual page 164

Logic analyzers
Hide thumbs Also See for 1660 Series:
Table of Contents

Advertisement

In the Data Memory Test menu, select Run, then select Single. The test runs one
17
17
time, then the screen shows the results. When the test is finished, select Done.
To run a test continuously, select Repetitive. Select Stop to halt a Repetitive Run.
To exit the tests, press the System key. Select the field to the right of the Sys PV
18
18
field.
Remove the disk containing the performance verification test, then select the Exit
19
19
Test System.
If you are performing the self-tests as part of the troubleshooting flowchart, return to the
flowchart.
Troubleshooting
To run the self-tests
5–21

Advertisement

Table of Contents
loading

This manual is also suitable for:

1660a series1660as series1660a1661a1662a1663a

Table of Contents