Calibrating and Adjusting
This chapter gives you instructions for calibrating and adjusting the logic analyzer.
Adjustments to the logic analyzer include adjusting the CRT monitor assembly.
To periodically verify the performance of the analyzer, refer to "Testing
Performance" in chapter 3.

Logic analyzer calibration

The logic analyzer circuitry of the HP 1660A-series and HP 1660AS-series Logic
Analyzers does not require an operational accuracy calibration. To test the logic
analyzer circuitry against specifications ( full calibration) , refer to chapter 3,
Testing Performance.
The oscilloscope circuitry in the HP 1660AS-series Logic Analyzers requires an
operational accuracy calibration by the user or service department under any of the
following conditions:
at six months intervals or every 1,000 hours
if the ambient temperature changes more than 10° C from the temperature at full
calibration
to optimize measurement accuracy
To test the oscilloscope circuitry against specifications ( full calibration) , refer to
chapter 3, Testing Performance.