c c Adjust the pulse generator channel 2 Delay, then select Precision Edge Find in the
oscilloscope Delta T menu. Repeat this step until the pulses are aligned according to
the setup time of the setup/hold combination selected, +0.0 ps or − 100 ps.
Select the clock to be tested.
4 4
a a Select the clock field to be tested, then select the clock as indicated in the table. The
first time through this test, use the top multiple-edge clock in the following table.
Clocks
J
K
L
M
N
P
b b Connect the clock to be tested to the pulse generator channel 2 output.
c c Select Done to exit the Master Clock menu.
To test the single-clock, multiple-edge, state acquisition (logic analyzer)
3–55