Set up the Format menu.
2 2
a a Press the Format key. Select State Acquisition Mode, then select Full Channel/4K
Memory/100MHz.
b b Select the field to the right of each pod field, then select ECL. The screen does not
show all pod fields at one time. Use the knob to access pod fields not shown on the
screen.
Set up the Trigger menu.
3 3
a a Press the Trigger key. Select Clear Trigger, then select All.
b b Select the Count Off field. Select Time in the pop-up menu, then select Done to exit
the menu.
To test the single-clock, multiple-edge, state acquisition (logic analyzer)
3–49