Set up the logic analyzer

Set up the Configuration menu.
1 1 1 1
a a Press the Config key. At the pop up menu, select Scope Channel.
b b Select the Input field, then select C1.
c c Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
d d Move the cursor to the V/Div field, then use the PRG knob to dial in 400 mV.
e e Move the cursor to the Offset field. Set the offset to 0 by typing 0, then pressing the
Select key.
f f Select the Coupling field, then select 50Ω / DC.
g g Move the cursor to the s/Div field, then use the RPG knob to dial in 2.00 ns.
Set up the Display menu.
2 2
a a Press the Display key.
b b Select the Mode/Arm field, then select Normal.
c c Select the Grid field and set it to On.
d d In the Waveform selection menu, delete channel 2. If channel 1 is not inserted, insert
channel 1.
To test the time measurement accuracy (oscilloscope)
3–87