To test the time interval accuracy (logic analyzer)
Set up the function generator according to the following table.
3 3
Function Generator Setup
Freq: 200 000 . 0 Hz
Amptd: 3.000 V
Phase: 0.0 deg
DC Offset: 0.0 V

Set up the logic analyzer

Set up the Configuration menu.
1 1
a a Press the Config key.
b b In the Configuration menu, assign Pod 1 to Machine 1. To assign Pod 1, select the
Pod 1 field, then select Machine 1.
c c In the Analyzer 1 box, select the Type field, then select Timing.
3–60
Main Function: Square wave