HP 1660 Series Service Manual page 163

Logic analyzers
Hide thumbs Also See for 1660 Series:
Table of Contents

Advertisement

Troubleshooting
To run the self-tests
Select Board Tests, then select Run. When the Board Tests are finished, select Done.
13
13
Select Data Input Inspection. All lines should show activity. Select Done to exit the
14
14
Data Input Inspection.
If you have an HP 1660A-series Logic Analyzer ( no oscilloscope) , go to step 18. If you have
an HP 1660AS-series Logic Analyzer, continue with step 15.
Select Analy PV, then select Scope PV in the pop up menu. Select Functional Tests.
15
15
Select one of the Scope PV tests.
16
16
You can run all of the tests at one time by selecting All Tests, or you can run each test
individually. For this example, select Data Memory Test.