HP 1660 Series Service Manual page 175

Logic analyzers
Hide thumbs Also See for 1660 Series:
Table of Contents

Advertisement

Troubleshooting
To test the logic analyzer probe cables
Using four 6-by-2 test connectors, four BNC Couplers, and four SMA (m) - BNC (f)
7 7
Adapters, connect the logic analyzer to the pulse generator channel outputs. To
make the test connectors, see chapter 3, "Testing Performance."
a a Connect the even-numbered channels of the lower byte of the pod under test to the
pulse generator channel 1 Output and J-clock.
b b Connect the odd-numbered channels of the lower byte of the pod under test to the
pulse generator channel 1 Output.
c c Connect the even-numbered channels of the upper byte of the pod under test and the
clock channel to the pulse generator channel 2 Output.
d d Connect the odd-numbered channels of the upper byte of the pod under test to the
pulse generator channel 2 Output.
On the logic analyzer, press Run. The display should look similar to the figure below.
8 8
If the display looks like the figure, then the cable passed the test.
9 9
If the display does not look similar to the figure, then there is a possible problem
with the cable or probe tip assembly. Causes for cable test failures include the
following:
open channel.
channel shorted to a neighboring channel.
channel shorted to either ground or a supply voltage.
Return to the troubleshooting flowchart.
5–32