2 S a f e t y a n d C a r e

Battery Pack Replacement and Disposal

Replace the battery pack when it is no longer able to hold a charge.
When the system is used with the Mobile QuikStart option and
frequently disconnected from the AC power (for a duration of
greater than four to eight hours), replace the battery every 12 to 18
months.
When the system is used with the Mobile QuikStart option and only
occasionally disconnected from the AC power (for a duration of
approximately four hours), replace the battery every three years.
Example of an AC Tray panel with the battery pack charge indicator LED
The battery pack charge indicator LED is located at the rear of the system,
on the AC Tray panel. A green blinking LED indicates that the battery is
actively charging. A solid green LED indicates that the battery pack is fully
charged. If the LED is not illuminated, then there is a problem with the
battery pack, or the battery pack may be missing.
2 - 2 4
1 Battery pack
charge
indicator LED
2 AC OK
indicator LED
3 Mains ON-OFF
circuit breaker
.
[ 1 ]
I N S T R U C T I O N S
F O R
U S E