Measurement Range and Accuracy
The following tables describe the variability in accuracy for clinical
measurements.
Clinical Measurements: Range and Accuracy
Direct
Measurement
Range
Distance*
0 – 36 cm
Distance using
0 – 60 cm
SieScape extended
field of view
Trace Area*
0 – 560 cm
Time @
0 – 9.4 sec
Slow Sweep
Time @
0 – 4.7 sec
Med Sweep
Time @
0 – 2.4 sec
Fast Sweep
Velocity
10 – 600 cm/sec
measurements on MPR (slice) quadrants from acquired volumes
* Also applies to
[ 1 ]
I N S T R U C T I O N S
F O R
Tolerance
Symbol
Tolerance Function or Value (Accuracy)
Tol(D)
3% of the distance or 1.5mm; whichever is greater
assuming 1540 m/sec speed of sound. Does not
apply to trace tool. Distance tolerance using trace
tool is user-dependent.
Tol(D)
Linear transducer: 5% of the distance or 1.5 mm;
whichever is greater assuming 1540 m/sec speed
of sound.
Curved transducer: 8% of the distance or 1.5 mm;
whichever is greater assuming 1540 m/sec speed
of sound.
2
Tol(A)
6% of the area or 1.5cm
assuming minimal operator error in tracing the
desired object and assuming 1540 m/sec speed of
sound.
Tol(T) @
15 msec
Slow Sweep
Tol(T) @
7.5 msec
Med Sweep
Tol(T) @
3.8 msec
Fast Sweep
Tol(V)
10% of the velocity or 5 cm/s; whichever is greater
using a J J & A string phantom at 45 degrees
±1 degree.
U S E
6 T e c h n i c a l D e s c r i p t i o n
2
; whichever is greater,
.
6 - 1 9