3 S y s t e m C o n t r o l s
Doppler
During Doppler, the group box contains drop-down menus for determining
the Update style of your 2D image and Doppler spectrum.
To enable the refresh format or change the refresh interval:
Select the refresh interval setting from the Trace Update Styles drop-
down list in the Imaging group box.
During update, the system pauses the Doppler spectrum and refreshes
the 2D-mode image according to the interval defined using the
Trace Update Styles drop-down list. Available settings are:
2D-Lv/D-Lv
2D-Lv/D-Frz
D-Lv/2D-Frz
D-Lv/2D-Lv
D-Lv/2D-EOS (end of sweep)
D-Lv/2D-2S (every two seconds)
D-Lv/2D-4S (every four seconds)
D-Lv/2D-8S (every eight seconds)
To toggle frozen and real-time images:
Press the UPDATE VIEW key.
To enable a simultaneous format:
Select 2D-Lv/D-Lv or D-Lv/2D-Lv from the Trace Update Styles drop-
down list in the Imaging group box.
Both the 2D-mode image and the Doppler spectrum display
simultaneously in real-time.
3 - 3 4
[2] Instructions for Use
IMAGE:
Doppler functions
CALCS:
Using the Auto
Statistics Tool